|
MI-600 Ca++
Ion Microelectrode
Operating
Instructions |
 |
Carefully unwind the tape and remove
the probe from the protective glass
tube. Only the bulbous tip is
sensitive to calcium ion. For best
results, soak the tip of the
electrode in 0.01 M CaCl 2 solution
for 24 hours prior to use.
The use of a
separate Reference Electrode is not
required.
Calibration
Meter requirement: Any research
grade pH meter with millivolt mode
capability. Set meter to millivolt
mode.
The
MI-600 is standardized using pure
Calcium Chloride (CaCl2) solutions
and again in solutions containing
possible interfering ions (if
appropriate). The pure CaCl2
solutions are used to determine
probe function. In pure solutions,
a 25-30 mv difference (approximate)
will occur between each tenfold
change in concentration. For
example, test in 0.001 M CaCl2, 0.01
M CaCl2, and 0.1 M CaCl2.
Standardization in solutions
containing possible interfering ions
is done in order to simulate the
actual samples to be analyzed. For
example, if your samples contain a
known sodium background such as 150
millimoles NaCl then your
calibrating standards should also
have this background.
These are some
possible mixtures you could use:
0.001M CaCl2-0.15 N NaCl, 0.01M
CaCl2-0.15 N NaCl, 0.1M CaCl2-0.15
N NaCl.
The
data are then plotted on semilog
paper (See Sample Calibration
Curve).
Remember! For best results, do the
following:
-
Always use standards that
bracket the concentration range
that you expect to have in your
samples.
-
Always mix your standards to
closely match the composition of
your samples.
-
Always calibrate using the same
volume of standard that you will
have as a sample.
Example: samples = 50 microliters
then standards = 50 microliters.
Handling
Always handle the electrode with the
same care you would use with other
glass electrodes.
Cleaning
When using the electrodes in
solutions containing protein, the
MI-600 and the reference electrode
should be rinsed with an enzyme
cleaning solution such as
Terg-a-zyme
(Alconox, Inc.) after each use
for a couple of minutes to remove
the protein from the membrane and
the reference junction. This will
prolong the useful life of the
electrodes.
Storage
Always clean the microelectrode
before storing:
Long-term (over 2 weeks):
Return the electrode to its original
container and prepare it in the same
condition is which you received it.
Usually this means simply returning
the electrode to its original
protective shipping tube after
cleaning.
Short-term: The
electrode can be left in a 0.01 M
CaCl2 solution.
|