MI-720, 8-720 and 16-720 CO2 Microelectrodes
Operating Instructions

The MI-720, 8-720 or the 16-720 Electrode Kit contain the following:

1   Carbon Dioxide Electrode with membrane housing
6   Replaceable membrane housings
1   Bottle of electrolyte
1  Syringe with Filling Tips
1  Set of Instructions

Electrode Assembly and Preparation

1.  Remove the acrylic housing from the Carbon Dioxide electrode by unscrewing it from the electrode body.  This housing, with affixed Teflon membrane, does not contain the necessary carbon dioxide electrolyte solution.

2.  Add Carbon Dioxide electrolyte solution to the acrylic housing.  This is accomplished by attaching a filling tip to the bottle of electrolyte and adding the electrolyte to the housing to a minimum height of 6 mm.  Gently place the bubble-free end of the filling tip against the Teflon membrane and release the electrolyte to the minimum height.

3.  Insert the housing on to the Carbon Dioxide electrode, being careful not to trap any air bubbles near the electrode tip.  The housing is screwed clockwise into the body of the electrode until it stops.  Check the tip of the electrode for proper seating of the housing against the electrode.  The electrode is properly seated if the electrode protrudes slightly beyond the end of the housing. 

Calibration

Calibration of the electrode requires the use of two standard gases with percent values that are close to the percent values of Carbon Dioxide to be measured.  Common values used are 5% carbon dioxide for zeroing and 10% Carbon Dioxide for sloping or gain.  A working calibration curve must be constructed using semilog graph paper.  Following the calibration procedures is an example of a working calibration curve.

A.  Calibration for gas samples:

When samples to be measured are gaseous, calibration should be performed with humidified gases.  Two possible setups for calibrating the electrode are shown below.  Keep the tip of the electrode as far as possible from the surface of the water.  The bubbling rate of the gas should be slow (3 - 6 bubbles per second).  Although  bubbling at a faster rate will flush the chamber more quickly, it will also cause a cooling effect on the electrode.

(Insert Picture)

1.  Bubble the 5% gas through the chamber and record the millivolt value after a stable reading is obtained.   It may take up to 15 minutes to de-gas the chamber of contaminants, however bubbling the gas vigorously will cause droplets to collect on the tip of the electrode.  This will make the electrode response time appear to be slow.

2.  Bubble the 10% gas (or any other percent value you decided to use for your application) through the chamber until a stable reading is obtained.  Record the reading.

This procedure of alternating between the two gases should be continued until you become confident of stability and reproducibility.  After creating a working calibration curve, the electrode is now ready to use.

B.  Calibration for liquid samples

To decrease calibration time, two separate calibration chambers should be used.  One for the 5% gas and another for the sloping gas such as 10%.  When setting up the calibration chambers initially, it will take up to 30 minutes to flush each chamber to obtain a steady state Carbon Dioxide level and a constant temperature.  Again the bubbling rate should be carefully regulated (3 - 6 bubbles per second) so that both calibrating liquids are at the same temperature.

(Insert Picture)

Calibrating standards and samples must be at the same temperature for accurate Carbon Dioxide measurements.

1.  Immerse the tip of the electrode into the 5% standard and record the millivolt value after a stable reading is obtained. after a stable reading is obtained.

2.  Remove the electrode from the first standard and place it into the second standard.  Record the reading after a stable reading is obtained.

Alternate between the two standards until you become confident of stability and reproducibility.  After creating a working calibration curve, the electrode is now ready to use.

Handling

When necessary, the membrane of the electrode can be replaced by following the assembly procedure above.  When removing and replacing a membrane as well as when calibrating or making measurements, be careful not to apply pressure against the internal electrode.  Any excessive pressure against the internal electrode can cause the electrode to crack rendering it useless and unrepairable.

Cleaning

When using the electrode in solutions containing protein, the electrode should be soaked in an enzyme cleaning solution such as Terg-a-zyme (Alconox, Inc.) after each use for a couple of minutes to remove the protein from the membrane surface.  This will prolong the useful life of the membrane.

Storing

Always clean and rinse the electrode before storing:

Long-term (over 1 week):  Remove the membrane housing from the electrode.  Rinse the internal electrode with distilled water and pat dry.  Place a new, unfilled membrane housing over the internal electrode and attach loosely (Do not seat completely).  This membrane will serve to keep the dust off of the electrode tip.

Short-term:  The sensing tip of the electrode with membrane housing still attached can be left soaking in a small amount of the Carbon Dioxide electrolyte.

Output Conversion

A.  Formula for conversion of percent Carbon Dioxide to solubility in moles/liter:

S= (a/22.414) x (760-p)/760) x (r%/100)

S= solubility of gas in moles per liter                  
a= absorption coefficient of gas at temperature
P= vapor pressure of water at temperature      
r%= actual reading in percent Carbon Dioxide                 

Temperature (T) vs. absorption coefficient (a) for oxygen in water:

T a T a
5 1.424 19 0.902
6 1.377 20 0.878
7 1.331 21 0.854
8 1.282 22 0.829
9 1.194 23 0.804
10 1.154 24 0.781
11 1.117 25 0.759
12 1.083 26 0.738
13 1.050 27 0.718
14 1.019 28 0.699
15 .03415 29 0.682
16 0.985 30 0.665
17 0.956 35 0.592
18 0.928 40 0.530

T = Degrees Celsius          
a = Absorption Coefficient

 

B.  Formula for conversion of percent Carbon Dioxide to mm of Hg (torr):

P = (r%/100)x(Pb-Pw)

P    = partial pressure of gas in mm of Hg (torr)
           r% = percent Carbon Dioxide (calibrating standard)
Pb = barometric pressure                                 
Pw  = water vapor pressure at temperature    

Sample Handling

The electrode measures dissolved Carbon Dioxide, therefore handling of your sample is critical for accurate measurement.  For blood gases, samples should be measured immediately minimizing exposure of the sample to room air.  For measuring total Carbon Dioxide, samples must be pH adjusted to a pH of 5 using a citrate buffer for conversion of carbonate and bicarbonate ions to dissolved Carbon Dioxide.

 

 

 

 