MI-403 Double Junction
Reference Microelectrodes
Operating Instructions
|
 |
The Reference Electrode contains the
following:
The MI-403 Double Junction Reference
microelectrode can be used with any
pH, ion-selective or redox
electrode. The electrode is
composed of an internal filling
solution of silver-silver chloride
electrode with an internal filling
solution of 3 M KCl saturated with
AgCl. The inner junction is
silicate fiber sealed in glass. The
outer junction is a ceramic frit.
Before the electrode can be placed
into operation, the internal glass
reference barrel must be filled with
the reference solution supplied.
The outer reference chamber must be
filled with a suitable electrolyte.
Filling the Reference Barrel
-
The glass tube is removed from
the electrode cap by grasping
each end of the cap and pulling
them apart. Remove the inner
glass reference barrel from the
outer chamber for easier
filling.
-
The internal reference solution
is added to the inner glass tube
using the polyethylene tubing
(filling fiber) provided.
-
After filling the inner glass
reference barrel with the
reference electrolyte, the inner
reference barrel is re-inserted
into the outer chamber, the
silver wire is inserted into the
glass tube and the electrode cap
is re-assembled.
-
Slide fill hole cover from
filling hole of outer chamber.
Fill the outer chamber with an
appropriate electrolyte solution
to just below the fill hole.
-
After initial filling, allow the
tip of the reference electrode
to soak in the outer filling
solution for 1 hour prior to
use.
Cleaning
When using the electrode in
solutions containing protein, the
electrode should be soaked in an
enzyme cleaning solution such as
Terg-a-zyme (Alconox,
Inc.) or a chromic/sulfuric acid
glass cleaning solution after each
use for 10 - 15 seconds to remove
the protein from the outer glass
barrel and the reference junction.
This will prolong the useful life of
the electrode. Storing
Storage
Always clean the microelectrode
before storing:
Long-term (over 4 weeks):
Remove the inner glass barrel
containing the internal electrolyte
and store the entire glass barrel in
a stoppered test tube filled with
the inner reference electrolyte.
Store the outer reference barrel in
an appropriate sized test tube
filled with the electrolyte used
inside the outer chamber. Rinse the
silver wire and electrode cap to
remove the salt solution and dry
using an absorbent towel. Store the
electrode in its original box or any
closed container to keep dust off of
the electrode.
Short-term: Place
the tip of the electrode into a test
tube or beaker containing the
reference electrolyte used in the
outer chamber. Our TC-10 Tissue
Culture Bottles with a B-10 Stopper
with hole is ideal for this purpose.
|