MI-404 Flat Membrane Microelectrodes
Operating Instructions
|
 |
The electrode is ready to use.
Carefully unwind the tape and remove
the probe from the protective glass
tube.
Calibration
The use of a
separate Reference Electrode such as
out MI-401 is required.
The MI-404 is
standardized in two pH buffers.
Follow the procedure recommended by
the manufacturer of your pH meter
for calibrating your pH meter with
our MI-404.
Optimum
Response Time: Optimum
response time will be obtained after
the probe has been exercised in two
buffer solutions. Place a pH 4.01
buffer or equivalent in a beaker and
a pH 6.86 buffer or equivalent in a
second beaker. Hold the MI-404 and
reference electrode together and
touch the pH 4.01 buffer allowing 15
- 20 seconds for equilibration. Do
this several times.
Handling
Always handle the electrode with the
same care you would use with other
glass electrodes.
Cleaning
When using the electrode in
solutions containing protein, the
electrode should be soaked in an
enzyme cleaning solution such as
Terg-a-zyme (Alconox,
Inc.) or a chromic/sulfuric acid
glass cleaning solution after each
use for 10 - 15 seconds to remove
the protein from the outer glass
barrel and the reference junction.
This will prolong the useful life of
the electrode.
Storage
Always clean the microelectrode
before storing:
Long-term (over 2 weeks):
Return the electrode to its original
container and prepare it in the same
condition is which you received it.
Usually this means simply moistening
the sponge located in the bottom of
the protective glass tube with
distilled water.
Short-term: The
electrode can be left in an acid pH
buffer solution, e.g. pH 4.01.
|