8-705 Flow-thru pH Electrode with 1/8" Fittings

8-705 Flow-thru pH Electrode with 1/8" Fittings

$299.00Price

8-705 Flow-thru pH Electrode with 1/8" Fittings

These electrodes are ideal for monitoring pH of a continuous flow such as water, urine, physiologic gluids, etc.

  • SPECIFICATIONS

    TOTAL LENGTH          

    37 mm tall

     

    2m  

    BODY (Outer Diameter)

    7 mm

    TIP (Outer Diameter)

    1/8" flow cell

    RESPONSE TIME:

    5 to 15 seconds 

    (dependent on flow rate, rinse...)

    swept volume 

    150 uL

    SENSITIVITY:

    pH 0 to 14  

    TEMPERATURE RANGE:

    -5°C to 100° C  

    REFERENCE ELECTRODES

    TYPE: Solution & Electrode

     

    3M KCI & Ag-AgCI

  • CALIBRATION

     

    The use of a separate Reference Electrode is required.

     

    The probe is usually used with the 8-702 reference electrode in series.

     

    Calibration is done by passing pH buffers through the two electrodes and then calibrating on the pH meter

  • CLEANING

    Using any electrode in solutions containing protein requires the electrode be rinsed with an enzyme cleaning solution.

     

    After each use, we recommend cleaning our electrodes with Terg-a-zyme (Alconox, Inc.) or a chromic/sulfuric acid glass cleaning solution by submerging the electrodes for a couple of minutes in order to remove all protein from the glass and reference junction.

     

    This will prolong the useful life of the electrodes.

  • STORAGE

    Always clean the microelectrode before storing:

     

    • Long-term (over 2 weeks): Return the probe to its original container and prepare it in the same condition in which you received it.  Usually this means simply moistening the sponge located in the bottom of the protective glass tube with pH 4 buffer.

     

    • Short-term: The probe can be left in an acid pH buffer solution (pH 4.01).